Chapter 4 - Setup

4-11

Error Return Values:

NOTALLOWED - the TareEnable parameter is
not set to true

“STATUSWORD” - means the command failed
because the weight was not stable (in motion) or
there was an A./D error.

SEQ

NUMBER

0

COMMAND#

(Hex)

1

STATUS

2

OFFSET

3

LENGTH

4

DATA

5-16

From M1 File

2

Write Non-Volatile
Command
(WRITENONVOLA
TILE)

The Write Non-Volatile Command - causes all param-
eters (including calibration constants) to be saved to
the non-volatile memory.

Error Return Values: None

SEQ

NUMBER

0

COMMAND#

(Hex)

1

STATUS

2

OFFSET

3

LENGTH

4

DATA

5-16

From M1 File

4

Reload Non-
Volatile
(RELOADNONVOL
ATILE)

The Reload Non-Volatile Command - causes the
weigh module to re-read the values stored in its non-
volatile memory. Any parameters changed since the
last write non-volatile command are overwritten. this
command can be used to abort a calibration or param-
eter entry session.

Error Return Values: None

SEQ

NUMBER

0

COMMAND#

(Hex)

1

STATUS

2

OFFSET

3

LENGTH

4

DATA

5-16

From M1 File

0x10