Cal low command (callowcmd), Cal high command (calhighcmd), C2 cal command (c2calcmd) – Hardy HI 1746-WS User Manual

Page 43

Advertising
background image

Chapter 4 - Setup

4-13

Cal Low Command
(CALLOWCMD)

The Cal Low Command - sets the “calLowCount”
parameter to the current A/D average counts when
doing a hard calibration.

Error Return Values:

STATUSWORD - there was a conversion error,
weight in motion or an A/D error or all three.

HARDCALFAILCOUNTS - there is less than
100 counts between the zero and the span
weights.

SEQ

NUMBER

0

COMMAND#

(Hex)

1

STATUS

2

OFFSET

3

LENGTH

4

DATA

5-16

From M1 File

0x64

Cal High
Command
(CALHIGHCMD)

The Cal High Command - Sets the Span Weight
parameter to the current A/D average counts when
doing Hard CAL.
Error Return Values:

STATUSWORD - there was a conversion error,
weight in motion or an A/D error or all three.

HARDCALFAILCOUNTS - there is less than
100 counts between the zero and the span
weights.

SEQ

NUMBER

0

COMMAND#

(Hex)

1

STATUS

2

OFFSET

3

LENGTH

4

DATA

5-16

From M1 File

0x65

C2 Cal Command
(C2CALCMD)

The C2 Cal Command - Performs a C2 Calibration.

Advertising