Setting the process weighing parameters, About the ladder logic example, Weight multiplier – Hardy HI 1746-WS User Manual

Page 50: Setting the process weighing parameters -20

Advertising
background image

HI 1746-WS WEIGH SCALE MODULE

4-20

12. M0:x.yy and M1:x.yyy (where x

= slot scanner module is in and
yyy = location within the M file)
are the M files being written to or
read from.

13. M0 is from the processor to the

scanner

14. M1 is from the scanner to the

processor

15. B11:0 is used for program control

bits.

Insure the following:

1.

Data file (i.e. N9) has 17 contigu-
ous words and the file number
and first word are listed in the
appropriate fields in the “COMM
SETUP” screen in WS100 and
the “Remote” box is checked.

2.

Proper driver is selected. This
should be the same driver that RS
Logix 500 would use.

3.

Station number of the processor
is correct which can be found in
the driver configuration.

4.

M file setup data (i.e. N10:0 to
N10:2) is correct for your setup.

5.

Slot location is correct in the M
file references.

Setting The Process
Weighing Parameters

About the Ladder
Logic Example

We have provided a Ladder Logic example explaining
how to set the weigh process parameters. The Ladder
Logic example is meant to provide a ladder logic
model only. Your application may vary and the exam-
ple may or may not meet your requirements.

Weight Multiplier

To use one of the commands, place the offset of a
parameter, from Table 4-1, into the command-
Data.offset of the M0file. If you are setting a parame-

Advertising