FLUKE 900 SERVICE MANUAL

Seiftest

Tests 0,47 and SO are not automatically run during the poweran selftest, but can be nin by
themselves explicitly. This is also the case for a group of tests known as "engineering" or

"extended" tests which are designated Exx as follows:

E60

WALK SIZELEDS

E61

TEST MEMORY

E62

WALK MONLEDS

E63

UPDATE_VCC_UM

E64

SHAD ADDR

E70

KEY CLOSURE

E71

DISP RAM

E74

DISP CHAR SET

E75

CART_SLFT

E76

TEST C ENG

E77

ACT_CROSS

E78

TRG CROSS

E79

FRQ_CROSS

E80

FLT_CROSS

2 Flag Results

Flags may ^pear on lines 3 and 4 of the general results screen. A 0 means "OK" and a 1
means "fault". At the present time, only the first row is used.

BYTEO

I

0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0

1 1 1 1 1 1 1 1

BIT: 76543210

BYTEl

BYTE2

I

I

0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0

BYTE3

I

0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0

BYTEO: Indicates DRAM chip which failed memory test

B7: U86
B6: U82
B5: U87
B4: U83
B3: U88
B2:U84
Bl: U89
BO: U85

3-3