Test 15 pullpos_freq, Tests 16,17, Trigqalo_equal, trigqal1_equal – Fluke 900 User Manual

Page 32: Tests 18,19, Qal0_equal, qal1_equal, Trigo_equal, trig1_equal, Tests 22 23, Pullo_trig_equal, pull1_trig_equal, Tests 22,23 pullo_trig_equal, pull1_trig_equal - 8, Test 24 pullo trigqal

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

Seiftest

FLUKE 900 SERVICE MANUAL

TEST 15

PULLPOS_FREQ

This test walks a positive pulse through the pull-ups latch to tiie frequency circuitry,

TESTS 16,17
TRIGQALO_EQUAL, TRIGQAL1_EQUAL

These tests walk a 0 or a 1 through the trigger and qualifier latches and the results are
determined through reading the EQUAL line.

TESTS 18,19

QAL0_EQUAL, QAL1_EQUAL

These tests walk a 0 or a 1 through the qualifier latches and the results are determined
through reading the EQUAL line.

NOTE:

For tests 18 and 19, an X in the general test result means that individual

results are imavailable.

TPQT^ Pn P1

TRIGO_EQUAL, TRIG1_EQUAL

These tests walk a 0 or a 1 through the trigger latches and the results are determined through
reading the EQUAL line.

NOTE:

For tests 20 and 21, an X in the general test result means that individual

results are unavailable.

TESTS 22 23

PULLO_TRIG_EQUAL, PULL1_TRIG_EQUAL

These tests walk a 0 or a 1 through the pull-ups latches, through the trigger circuitry and are
read through the EQUAL line.

NOTE:

For tests 22 and 23, an X in the general test result means that individual
results are unavailable.

TEST 24

PULLO TRIGQAL

This test walks a 0 through the pull-tq>s latches, through the trigger and qualifier circuitry and
the results are determined through reading the EQUAL line.

3-8

Advertising