FLUKE 900 SERVICE MANUAL

Maintenance

TABLE 1

Offset Shift

Fluke P/N

R

4 0

80

120 160 200 240

365k

0

0

0

1

1

1

921168

150k

1

1

2

2

3

3

921171

113k

1

2

2

3

4

4

921176

8 6k

2

2

3

4

5

5

921184

57k

3

4

5

6

6

7

921189

47k

4

5

6

6

7

8

921192

TABLE 2

Offset Shift

Fluke P/N

C

40

80

120 160 200 240

1. OpF

1

1

1

2

2

2

921197

1.5pF

1

1

2

2

3

3

921200

1.8pF

1

1

2

3

3

3

921205

2.2pF

2

2

3

3

4

4

921213

3.3pF

2

3

3

4

4

5

921218

4.7pF

2

3

4

5

6

6

921221

6.8pF

3

4

6

6

7

8

921226

NOTE: All the offsets in Table 2 are meant as negative (i.e. they must be subtracted from
the values in the calibration data printout).