FLUKE 900 SERVICE MANUAL

Selftest

3.4 System Error Codes

System Error Codes appear on the command line of the 900 when its microprocessor operation

hangs up. These codes are used mainly for design debug, but are included here for completeness
in the event a software bug is encoimtered. The most common cause of such a system crash is the
loading of a corrupted simulated reference device file.

These codes are not of much use for hardware troubleshooting. The only hardware code is 80 and
its message spears similar to the following:

System Error #80

PC = 536A
Overlay = Selitest
Cannot clear Fault Latches

This indicates that at the time of the crash, the microprocessor Program Counter contained the
value 536A and memory addressing was in the Selftest overlay.

File Management Error Codes

01

FILE IS OPEN

02

DEVICE ALREADY FORMATTED

03

FILE NOT FOUND

04

FILE ALREADY EXISTS

05

INSUFHCIENT SPACE ON DESTINATION DEVICE

07

WRITTEN BYTE FAILED TO VERIFY

08

CANNOT WRITE DEVICE

09

FILE

CANNOT BE DELETED

OA DEVICE DOES NOT EXIST
OB FILE NOT OPEN
OC ERROR SETTING PAGE
OD ERROR SETTING CHAPTER
OE ERROR ACCESSING DEVICE
OF PASSWORD INVALID

10

INSUFnCIENT STACK SPACE FOR TRANSFER

11

UNABLE TO COMPRESS

12

CHECKSUM ERROR ON FILE

13

COPY PROTECTED

14

DEVICE NOT FORMATTED

15

FILE

UNRECOVERABLE/NOT FOUND

16

FILE IS DELETE PROTECTED

17

DEVICE HAS NO OVERHEAD

18

DIRECTORY EXCEEDS PAGE OVERHEAD

19

DCB DATA DOES NOT MATCH DEVICE DATA

lA ERROR IN OVERHEAD DATA
IB ADDRESS IS OUT OF FILE BOUNDS
1C FILE IS TOO BIG TO BE LOADED
ID DEVICE PASSWORD INCORRECT
IE END OF FILE REACHED

3 - 2 7