FLUKE 900 SERVICE MANUAL

Selftest

NOTE:

For test 24, an X in the goieral test result means that individual results are
unavailable.

TESTS 25,26

FM_STAT_NFLT, FM_STAT_FLT

These tests perform a static FMASK test All lines are expected to pass for the "_NFLT" test

while all lines are expected to fail for the "FLT" test

TEST 27

FMASK_WB_TEST

This test does an approximate test using fixed values for FMASK. It uses a tolerance three
times that of the FMxx tests. This test is performed to determine whether or not the FMASK
curve is within a very wide margin.

TEST 28

FMASK_CAL

This routine determines the calibration values to be used by FMASK.

TESTS 29,31,33,35,37,39

FMxx_FLT

These tests use a pulse of fixed duration (40, 80,120,160

,200

and 240ns) and set FMASK to

just below the duration of the pulse. All lines are expected to produce a fault Those which

do not are flagged as faulty.

TESTS 30,32,34,36,38,40

FMxx_NFLT

These tests use a pulse of fixed duration (40, 80,120,160,200 and 240ns) and set FMASK to

just above the duration of the pulse. AH lines are expected to produce no fault Those which

do are flagged as faulty.

TEST 41
ACT_DIS_CLR

This test checks the disabling and clearing functions of the activity circuitry. It disables the

activity circuit and pulses all lines. Any lines which indicate activity are flagged as faulty. It
also verifies that the activity latches may be cleared and that a single edge on all lines

produces a 1 (activity occurred) in all latches.

3-9