Maintenance

FLUKE 900 SERVICE MANUAL

5.4.2.6.4

LM360 comparator related condition

Each pair of columns in the calibration results should differ by no more than 1. As an
example:

. . 1 6 0

n s . .

pin# U-H D-H

U-L D-L

5

-

1

0

+2 +2

Check that (U-H) - (D-H) = Oor +/-1 and

(U-L) -(D-L) = Oor+/-1

If not, the LM360 comparator on this line is faulty and should be replaced. The
calibration procedure should be restarted from the beginning.

5.5 Board Revisions, Upgrades and ECOs

This manual documents aU 900 units with serial number 4720000 and above. Units with lower

serial numbers are documented in a two volume Service Manual (part numbers 859413, 880195).

The boards in such machines are designated IB, MB, Ml, M2 and HI.

The table below lists the boards and ECOs applicable to the 900 revision level described in this
manual.

BOARD

REVISION

PCB#

APPLICABLE ECOs

Interface Buffer

(IB)

300-506C

none

High Speed Board

(HS2)

300-508-2

HS2 has factory-installed ECOs

which make it equivalent to HS3

High Speed Board

(HS3)

300-508-3

none

Micro Board

(MB2)

300-507-2

ECO-016, ECO-018

5.5.1 Modification to MB for 64k Cartridges

ECO-016
This connects the 16th address line to the cartridge port.

Install wire from U9 pin 17 to J8 pin 5.

5 - 1 6