8.2.30 Zero Offset Adjustment

Zero o№et can be easily applied to make the flowmeter outputs comparable to process values
measured by other instruments.

If the zero adjustment described in “0 Still Water Zero Adjustment” can be performed, this zero offset

adjustment is not needed.

To change the zero offset value:

Calculate the zero offset value with the following equation:

Zero offset value (%) = {(actual flow rate) - (LF600F measured value)}

* The zero offset value should be calculated in percent value for Range 1 of converter. See the

following example.

(Example)

Measured condition

Flow rate

% in measuring span

Actual flow rate obtained

from other instrument.

10.0 mVmin

50%

LF600F measured value

10.5 mVmin

52.5 %

Zero offset

-2.5 %

(If zero offset is set to -2.5 %, the converter will output 50.0 % flow rate instead of -2.5%.)

-