2 output status for errors and alarms, Error or alarm message, Data display – Toshiba LF600F User Manual

Page 143: Current output (4^20ma), Totalizer and pulse output, Remarks, Toshiba

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6,F,8,A0,8,6,9,

12.2 Output Status for Errors and Alarms

The flowmeter data display, current and pulse outputs will become as follows if an error or alarm
occurs.

Error or alarm

message

Data display

Current output

(4^20mA)

Totalizer and

pulse output

Remarks

ROM ERROR

(Note 1)

(Note 3)

Stopped

After power-up, no

measurement starts.

RAM ERROR

(Note 3)

Stopped

PARAMETER FAIL

(Note 2)__________

Zero

(Note 3)

Stopped

EX. CURR OPEN

Zero

(Note 3)

Stopped

Zero adjustment

(on-stream at zero flow rate)

cannot be conducted.

EX. CURR ERROR

Zero

(Note 3)

Stopped

Zero adjustment

(on-stream at zero flow rate)

cannot be conducted.

ADC. ERROR

Zero

(Note 3)

Stopped

Zero adjustment

(on-stream at zero flow rate)

cannot be conducted.

INVALID TOTAL

Measured data

Measured data

Measured

data

The error message

disappears if you clear

(reset) the totalizer.

HIGH ALARM

Measured data

Measured data

Measured

data

LOW ALARM

Measured data

Measured data

Measured

data

HIGH HIGH ALARM

Measured data

Measured data

Measured

data

LOW LOW ALARM

Measured data

Measured data

Measured

data

Notes

1. The display and output may not be as indicated depending on the nature of the ROM error.

2. If a parameter failure relating to current output occurs, the current output may not become

exactly the setting value of the current output when an alarm occurs.

3. The current output set value used in case an alarm occurs will be output. For setting method,

see 0, “Current Output Setting Used When an Alarm Occurs.”

-

142

-

Advertising
This manual is related to the following products: