9.2.2 50 % Flow Rate Calibration

Using the converter's internal calibration circuit, 50% flow rate calibration can be performed. For the

calibration procedure, refer to the calibration procedure for 0% flow rate calibration (Or select the

"FLOW CAL 50" from the menu).

9.2.3100 % Flow Rate (Span) Calibration

Using the converter's internal calibration circuit, 100% flow rate calibration can be performed. For the
calibration procedure, refer to the calibration procedure for 0% (Or select the "FLOW CAL 100" from

the menu.).

9.2.4 Checking the Excitation Current Value

You can monitor the exciting current value.

■ To check the exciting current value:

Key operation

Display example

Description

0:

MEAS MODE
FLOW CAL 0
FLOW CAL 50
FLOW CALI 00

^EXCURR DSP

iSELl [ENTl

Select "EX CURR DSP" in the configuration item

selection screen.

[ENT]

04;

EX CURR DSP

0.2000

A

The excitation current value is displayed.

Pressing the [EXT] key returns you to the menu

screen.

_________ lEXJ]_________

The excitation current value is factory adjusted when shipped. Contact you're nearest Toshiba

representative if any change is necessary.

-