9.2 Calibration Using Converter Signal Source

9.2.1 0 % Flow Rate Calibration (zero point calibration)

Using the converter’s internal calibration circuit, 0% flow rate (hereafter called zero point)
calibration can be made.

To check the zero point of flow measurement:

Key operation

Display example

Description

0:

MEAS MODE

^FLOWCAL 0

FLOW CAL 50
FLOW CALI 00
EXCURR DSP

[SEL] [ENT]

Select "FLOW CAL 0" in the configuration setting
item selection screen.

STEP1

[ENT]

01:

FLOW CAL 0

0.1

%

[EXT] [SET]

The zero point by simulation input is displayed.

Then press the [SET] key.

* Pressing the [EXT] key returns you to the menu.

STEP2

[SET]

01:

ADJUST READY

0.1

%

When you press the [SET] switch, the title display

changes to ADJUST READY and enables you to

perform calibration.

* Pressing the [EXT] key returns you to the previous

[EXT] [SET]

screen.

STEPS

[SETjHold down

01:

FLOW CAL 0

*CAL 0% ADJ

Holdino down the [SET] kev when ADJUST READY

is displayed, "*CAL 0% ADJ" is displayed and

zero Doint calibration starts.

Switch operation is disabled during calibration.

STEP4

01:

FLOW CAL 0

0.0

%

After approximately 4 seconds, zero point

calibration is completed and a new zero point is

displayed.

Press the [EXT] key. You return to the menu

[EXT] [SET]

screen.

Note l:To start calibration, hold down the [SET] key.

Note 2: To cancel the adjustment when ADJUST READY is displayed, press the [EXT] key.

You return to the screen displaying the zero point by simulation input.

-