RATE CALIBRATION

YOU DON'T HAVE TO DO THIS.

Software is set to expect STANDARD hardware. If a device meters at a
much lower feed rate then expected, the software will take 10 to 20
cycles to fully adjust. During this time cycles will take longer.

An example of this is a system using a 1/2" auger instead of 1".

If you don't mind letting the system SELF ADJUST, or
your hardware is STANDARD,
then:
+---------------------------------------------------+

¦ ¦
¦ PROCEED TO: ENTERING SETTINGS NEXT PAGE ¦
¦ ¦
+---------------------------------------------------+

To perform a RATE CALIBRATION:

Have the HOPPER LOADED with enough material to run for several cycles
without running out.

Have the "OFF AT END OF CYCLE" and "PAUSE" switches DOWN.

Press: * Display will say: (PASSWORD)
Press: 22222 Display will say: (P x)

If you are calibrating an auger feeder, operate it briefly to ensure it
is fully primed. To do so:

Press: OPER Display will say: (OPERATE )

Press: 5 (or 6) Run until auger is dispensing.
Press: DUMP This will empty the weigh bin.

Now you can CALIBRATE the material.
To do so:

Press: CAL Display will say: (CALIBRATE)
Press: 5 Component #5 will calibrate itself.
(or whatever component you are calibrating)

Repeat this two key calibration routine for EACH material that you
wish to calibrate. Only components that have a TYPE selected (not
"OFF") will operate.

Each time a dispense will occur, followed by weighing, followed by
a dump to empty the weigh bin.

Press: EXIT when all calibrations are done.

YOU ARE NOW READY FOR ACCURATE PRODUCTION BLENDING.

+---------------------------------------------------+