Tighten the fillister head screws (L) of the

connection fasteners (Q) with a Torx driver

T20.

Close the hinged covers.

Replace diaphragms and valves of the opposite side of

the pump in the same way.

Feed the PTFE hose into the inlet of the

exhaust vapor condenser. Mount the con-

denser with the counter holder (Torx screws

T10) to the pump. Fasten union nut.

Check the overpressure safety relief device

at the exhaust waste vapor condenser.

If necessary, pull off the old pressure relief

valve and install the new one. Check for

correct position of the PTFE-foil under the

pressure relief valve.

Assemble catchpots (9) with joint clips.