Fault

Possible cause

Remedy

❑ Pump does not

achieve its ultimate

vacuum or usual

pumping speed.

➨ Pump too hot?

✔ Allow pump to cool down.

Determine and eliminate

the cause of overheating.

➨ Deposits have been

formed inside the

pump?

✔ Clean and inspect the

pump heads.

➨ Diaphragms or valves

damaged?

✔ Replace diaphragms and/

or valves.

➨ Outgassing substances

or vapor generated in

the process?

✔ Check process parame-

ters.

❑ Pump too noisy.

➨ Atmospheric or high

pressure at the pump

inlet?

✔ Connect hose or silencer

to pump outlet. Be careful

not to cause outlet over-

pressure, especially with

condensable vapors.

➨ Diaphragm crack or

diaphragm clamping

disc loose?

✔ Perform maintenance.

➨ Other than above men-

tioned causes?

✔ Contact local distributor.

❑ Pump seized.

✔ Contact local distributor.

A service manual with exploded view drawings, spare parts list and

directions for repair is available on request.

+

The service manual is intended for trained service people only.