Condensate and deposits will affect the measurement re-

sults. Clean the pressure transducer, if necessary. See

section “Cleaning the pressure transducer”, pg. 72.

Assembling the second CVC 3000 controller

(only PC 520 NT / PC 620 NT):

The pumping unit’s second CVC 3000 controller controls

the in-line valve at the second inlet (upper vacuum con-

nection). Mount the controller prior to operating the pump-

ing unit.

Affix the controller with its holding plate to

the pumping unit (three Allen screws (f), 2.5

mm wide Allen key). Install the washers and

screw the holding plate into position.

Slip the enclosed molded PTFE hose onto

the hose connection of the valve block at

the inlet (a). Fasten with hose clip. Close

hose clip with flat pliers.

Slip the other end of the PTFE hose onto

the vacuum connection (c) of the CVC 3000

and fasten with union nut.

Plug in the VACUU•BUS lines at the rear of

the controller (b): Power supply line of the

controller (d) and control line of the in-line

valve (e).

Attention: Do not apply off-axis forces when

assembling or removing plug connections!

Observe correct orientation of the plug.

Carrying the pumping units:

Hold the pumping unit with one hand at its handle and tilt

it slightly backwards.

Reach with the other hand underneath the pumping unit

and lift the pumping unit.

Or: Use the recessed grips on the side of the pump (below

pump heads). Do not grab if the pump is hot.

Attention: Do not grab the pumping unit at the holders of

the waste vapor condenser or of the catchpot!