To reduce pump noise emanating from the pump exhaust

port, connect an exhaust hose or use a silencer (see “Ac-

cessories”, pg. 86).

The VACUUBRAND controller CVC 3000 can only be oper-

ated with components compatible with the VACUUBRAND

VACUU•BUS system, (see “Accessories”, pg. 86).

The vacuum controller CVC 3000 controls VACUUBRAND

diaphragm pumps NT and optional in-line, coolant, and

venting valves. Connected components (e.g., venting

valve, level sensor, external pressure transducer 3000 se-

ries) are automatically identified and configurated. Identi-

cal components must be configurated beforehand; infor-

mation upon request.

Do not use more than one controller within the same

VACUU•BUS system. Several controllers in the same

VACUU•BUS system will interfere with each other and

result in error messages of the connected compo-

nents (pumps, valves).

CVC 3000

The CVC 3000 is equipped with an internal capacitive

pressure transducer with ceramic diaphragm. It measures

the actual pressure independently of the gas type, and

with reference to the vacuum, i.e., absolute.

Inside a vacuum system where evaporation occurs, e.g.,

a rotary evaporator, the vacuum is not uniform. For ex-

ample, a condenser can act as a pump, or the vacuum in

the connecting tubing can be higher or lower than in the

application itself. This affects the measurement results as

well as the control levels. Therefore, carefully choose the

position where to connect the vacuum controller sensor.

Position the controller CVC 3000 and its vacuum line in

such a way that condensate cannot flow towards the pres-

sure transducer.