Delay: Delay” determines the time the pump (with VMS module and

in-line valve) remains running and the coolant valve remains open af-

ter the process has been stopped.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that when the process stops, the pump is switched

off and the coolant valve closes immediately.).

+

Duration: Duration” determines the total process time since control

start.

A preset ”Duration” (process time) has no effect if the process is

stopped due to a preset ”Maximum” before ”Duration” is reached.

The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be

set to ”Off”. ”Off” means that no endpoint of the process is defined.

The screen-shot shows the factory-set values.

Vac control

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.
+

Press the selection knob twice to return to the standard display.

Vac control
Set vacuum 75 Torr

Hysteresis Auto

Maximum

Off

Delay

Off

Duration Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

760 Torr