VACUUBRAND PC 620 NT plastic housing User Manual

Page 89

Advertising
background image

page 89 of 113

Fault

Possible cause

Remedy

❑ Digital pressure

reading is flashing,

one blip*.

➨ Overpressure at the

pressure transducer,

pressure > 795 Torr

(1060 mbar)?

✔ Release pressure imme-

diately (risk of bursting).

❑ Warning triangle

and black valve

symbol are flash-

ing, two blips*.

➨ External venting valve

removed or defective?

✔ Connect valve or replace

with a new one or recon-

figure without valve.

❑ Warning triangle

and valve symbol

are flashing, three

blips*.

➨ NT VARIO / VARIO-B

pump and in-line valve

connected?

✔ Disconnect in-line valve;

switch controller off/on to

reconfigure.

❑ Warning triangle

and coolant valve

symbol are flash-

ing, four blips*.

➨ Coolant valve removed

or defective?

✔ Check connection cable

of the valve; or use new

valve or reconfigure with-

out valve.

❑ No digital pressure

reading. Warning

triangle is flashing,

five or seven blips*.

➨ Five blips: External

sensor defective or

removed?

✔ Plug in external sensor or

use new one or reconfig-

ure without sensor.

➨ Seven blips: Internal

sensor defective?

✔ Contact local distributor.

❑ Warning triangle

and pump symbol

are flashing, six

blips*.

➨ NT VARIO / VARIO-

B pump and VMS**

(Vacuum Management

System) connected?

✔ Remove one component.

Restart controller.

➨ Fault at the NT VARIO

pump or at the VMS?

✔ Check pump and/or VMS

or reconfigure.

➨ Fault at the NT VARIO /

VARIO-B pump?

✔ Check pump, restart con-

troller.

❑ Clock symbol is

flashing.

➨ Preselected process

time is over?

✔ Confirm by pressing

START/STOP key.

❑ Venting valve does

not respond, valve

symbol is displayed.

➨ External pressure

transducer connected

and active?

✔ Select internal sensor or

connect an external vent-

ing valve.

❑ Venting valve does

not respond, valve

symbol is dis-

played.

➨ Venting valve contami-

nated?

✔ Clean valve.

❑ ”Vac control” func-

tion: Control stops,

”arrow up” is flash-

ing.

➨ Preset maximum pres-

sure exceeded?

✔ Confirm by pressing

START/STOP key.

Change maximum pres-

sure value if necessary.

Advertising