Notice – VACUUBRAND PC 620 NT plastic housing User Manual

Page 21

Advertising
background image

page 21 of 113

cess will start immediately after a power failure with-

out pressing any further key. It is your responsibility to

ensure that automatic start-up of the system will not

lead to any dangerous condition. Provide appropriate

safety measures. Check

prior to starting the process

whether the option ”Autostart” (menu: configuration) is

enabled.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.

Adopt suitable measures to prevent any danger aris-

ing from hot surfaces or electric sparks. Ensure that

hot surfaces of the pump do not cause burns. Provide

a suitable contact guard if necessary.

• Ensure that the

coolant outlet pipe at the waste vapor

condenser is always free and that it cannot get blocked.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Check the liquid level in the catchpots regularly and drain

condensate in time to prevent overfilling. Install a level

sensor (see ”Accessories”, pg. 86) for monitoring, if

necessary.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

The CVC 3000 controller is powered by a short circuit

proof wide-range power supply with integrated overload

protection.

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Note: Only manual reset is possible. Switch off the pump

and disconnect from the power source. Identify and elimi-

nate the cause of failure. Wait approx. five minutes before

restarting the pump.