pots in time to avoid overflow. Install a level sensor (see

“Accessories”, pg. 86) for monitoring, if necessary.

The maximum liquid level is at approximately 80% of the

total filling level to avoid problems when removing the

catchpots.

Permissible range of coolant temperature at the exhaust

waste vapor condenser:

5 °F to 68 °F (-15°C to +20°C)

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Removing the catchpots:

Catchpot at outlet:

Remove joint clip. Remove catchpot

and drain conden-

sate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to restore at-

mospheric pressure in the catchpot before attempting

removal. Remove joint clip. Remove catchpot and drain

condensate.

Reattach drained catchpots.

+

Important: Comply with regulations when disposing

of solvents/condensates. Recycle if possible; purify if

contaminated.

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.