Controller

CVC 3000

Pressure transducer

ceramic diaphragm (alumina),

capacitive, absolute pressure,

gas type independent

Display

LCD graphic display, illuminated

Pressure units / scale (selectable)

Torr, mbar or hPa

Measuring range (absolute)

810 - 0.1 Torr (1080 - 0.1 mbar)

Maximum control range (absolute)*

795 - 0.1 Torr (1060 - 0.1 mbar)

Resolution

0.07 Torr (0.1 mbar)

Maximum permissible pressure at pres-

sure transducer (absolute)

1125 Torr (1.5 bar)

Maximum permissible temperature of

gaseous media**

continuous operation: 104°F (40°C),

for short periods (less than 5 minutes)

up to 176°F (80°C)

Measurement uncertainty (absolute)

after careful adjustment and at constant

temperature

<

± 0.75 Torr (1 mbar)

Temperature coefficient

<

± 0.05 Torr/K (0.07 mbar/K)

Ambient temperature range (operation)

50°F to 104°F (10°C to +40°C)

Ambient temperature range (storage)

14°F to 158°F (-10°C to +70°C)

Permissible relative atmospheric mois-

ture during operation (no condensation)

30% to 85%

Maximum permitted current of connect-

ed valves (connected components)

4A

Degree of protection IEC 529

(front side)

IP 42

Venting connection

hose nozzle for hose I.D. 3/16“ (4-5 mm)

Maximum admissible pressure at

venting connection

17.4 psi (1.2 bar) absolute

Interface

RS-232 C

* The actual vacuum control range in your application might be reduced due to ultimate vacuum

of the pump, volume of gas present, etc.

** if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)