During operation – VACUUBRAND PC 620 NT plastic housing User Manual

Page 41

Advertising
background image

page 41 of 113

• Ensure that the

coolant outlet tubing is always unob-

structed and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute. Outlet

flow must always be unhindered.

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or squeezed coolant hoses).

• Only install the optional coolant valve in the supply line

of the exhaust waste vapor condenser.

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

+

Maximum ambient temperature: 104 °F (40 °C)

Check the maximum temperatures, if installing the

pump in a cabinet or a housing. Make sure ventilation

is adequate, especially if the ambient temperature is

elevated.

+

Operation with silencer at the outlet: Operating the

pump at a high inlet pressure or pumping dusty gas-

es for a long time may cause clogging of the silencer.

Check the silencer regularly and replace if necessary.

• If the pump is installed at an altitude of more than 6500

ft (2000 m) above mean sea level, check compatibility

with applicable safety requirements, and adopt suit-