background image

page 95 of 113

(b)

(f)

(c)

(f)

PC 520 NT / PC 620 NT:

Remove the second controller and its holding plate.

Unplug in the VACUU•BUS lines at the rear

of the controller (b): Power supply line of

the controller and control line of the in-line

valve.

Attention: Do not apply off-axis forc-

es when removing plug connections!

Loosen the union nut at the vacuum con-

nection (c) of the controller. Remove the

PTFE hose from the vacuum connection.

Remove the three Allen screws (f) affixing the holding plate with a 2.5

mm wide Allen key. Remove the controller together with its holding

plate. Pay attention to the washers under the screws.

For maintenance, lay the pumping unit on its side with the pump heads

to be maintained at the top. Support the pumping unit appropriately.

Remove the 4 screws affixing the head cov-

er cowling (W) with a Torx driver T20. Pay

attention to the washers under the screws

and remove.

Pull off head cover cowling (W) carefully. Do

not tilt.