• The equipment is designated for an ambient and gas inlet temperature

during operation of +10 to +40°C. Never exceed these ambient and

gas inlet temperatures. If pumping / measuring gases which are not

potentially explosive, extended gas inlet temperatures are permissible.

See instructions for use, section “Gas inlet temperatures” or “Techni-

cal data”.

After any intervention at the equipment (e.g., repair / maintenance) the ul-

timate vacuum of the pump has to be checked. Only if the pump achieves

its specified ultimate vacuum is the pump’s leak rate low enough to en-

sure that no explosive atmospheres will occur in the interior of the equip-

ment.

After any intervention at the vacuum sensor, the leak rate of the equip-

ment has to be checked.

Attention: This manual is not available in all

languages of the EU. The user must not op-

erate the device if he does not understand

this manual. In this case a technically cor-

rect translation of the complete manual has

to be available. The manual must be com-

pletely read and understood before opera-

tion of the device. All specified measures

must be applied, or else must be replaced

by equivalent measures at the user’s own

risk.