Fault

Possible cause

Remedy

❑ ”Pump down” func-

tion: Control stops,

”arrow down” is

flashing.

➨ Pressure below preset

minimum pressure?

✔ Confirm by pressing

START/STOP key.

Change minimum pres-

sure value if necessary.

❑ No function is dis-

played, no menu is

available.

➨ No controllable device

(valve, VMS, VARIO

pump) connected?

✔ Connect device at CVC

3000. Switch CVC 3000

off/on to reconfigure.

❑ Controller does

not respond when

pressing keys (ex-

cept ON/OFF).

PC symbol is dis-

played.

➨ Controller in remote

mode?

✔ Control CVC 3000 via in-

terface or switch off re-

mote mode.

❑ Controller does

not respond when

operating any keys.

No change after

switching off/on.

✔ Contact local distributor.

❑ Pump does not

start or stops im-

mediately. Warning

triangle and pump

symbol are flash-

ing.

➨ Motor overloaded?

✔ Allow motor to cool down,

identify and eliminate

cause of failure. Manual

reset is necessary. Switch

off pump or unplug.

➨ Overpressure in outlet

line?

✔ Remove blockage in line,

open valve.

➨ Pump has been ex-

posed to condensate?

✔ Allow pump to run for

some minutes with atmo-

spheric pressure at the

inlet.

❑ Pump does not

achieve its ultimate

vacuum or usual

pumping speed.

➨ Leak in the pipeline or

in the vacuum system?

✔ Check connections, pipe-

line and vacuum system.

➨ Wrong setting at con-

troller?

✔ Select function ”Pump

down” and check again.

➨ Long, narrow vacuum

line?

✔ Use lines with larger di-

ameter, length as short as

possible.

➨ Pump has been ex-

posed to condensate?

✔ Allow pump to run for

some minutes with atmo-

spheric pressure at the

inlet to purge.