background image

5. Wéighing

Cautions for the weighing operation

□ Operate your balance gently.

Press the RE-ZERO key to prevent possible error before placing material

on the pan (weighing material) each time.

Shorten the operation time as much as possible. ( Opening and closing
door, placing and removing material)

Temperature changes during measurement may cause weighing error.

Use a pair of tweezers (pincette) to avoid a temperature change that Is due

to having your hand in the weighing chamber.

Calibrate your balance periodically to maintain weighing accuracy. Refer
to section "T.Calibration".

Electrified material or magnetic body may cause a weighing error.

Do not press keys with a sharp instrument (such as a pencil or ball
point pen).

Do not drop things on the pan, or place a weight on the pan that is

beyond the weighing range of the balance.

Keep the area clean and dry.

Consider section "2. Caution" for weighing operation.

Basic Operation (gram mode)

Step 1 Calibrate your balance before use. (Refer to section "7. Calibration")

Step 2 Place a container on the weighing pan, if necessary.

Press the | RE-ZERO | key to cancel net weight. The balance displays zero.

Container:

A vessel placed on the pan, but not to be included in the

weighing data.

Step 3 Place material on the pan or in the container.

Step 4 Wait for the stabilization indicator to be disipayed and read the value.

Step 5 Remove the material and container from the pan.

Basic operation

page 13