Cautions during use (To get best performance)

Note the following items to get accurate weighing data.

Press the | RE-ZERO | key before each weighing to prevent

possible error.

Calibrate the balance periodically so as to cancel possible

weighing error.

Make each weighing quickly to avoid errors due to changes

in the environmental conditions.

Close glass doors to keep out drafts.

Do not drop things upon the weighing pan, or place a

weight beyond the range of the balance on the weighing

pan.

Do not use a sharp instmment (such as a pencil or ball point
pen) to press the keys, use your finger only.

Toweighproperlywithaminimum display of 0.01 mg using

the GR-202, replace the breeze break ring with the fine

range breeze break ring. See page 8 for details.

Dischargestaticelectricityfromtheweighed matter. When

weighing material (plastics, insulator, etc.) could have a

static charge, the weight value is influenced. Try to keep

the ambient humidity above 45%RH or to use the metal

shield case.

This balance uses a strong magnet as part of the balance

assembly, so please use caution when weighing magnetic

materials. If there is a problem, use the underhook (on the
bottom of the balance) to suspend the material away from

the influence of the magnet.

Cancel the temperature difference between the weighed

material and the environment. When a sample is warmer

(cooler) than the ambient temperature, the sample will

lighter (heavier) than tme weight. This error is due to the

rising (falling) draft next the sample.

Take into consideration the affect of air buoyancy on a

sample when more accuracy is required.