Calibration Using the Internal Weight

This function calibrates the balance using the internal weight.

Operation

Step 1 Connect the AC adapter and warm up the balance for

at least one hour with nothing on the weighing pan.

Step 2 Press the CAL| key to start calibration.

Step 3 The balance displays

\ [ n j ~ m \

and performs calibra­

tion. Prevent vibration and drafts from affecting the

balance.

Step 4 Ifthe "CLP output ( ,n/-o)" of the "Function Table" isset

to
232C interface.

or "P", "Calibration Report" is output from RS-

Step 5 The balance will automatically return to the weighing

mode after calibration.

Step 6 Test the accuracy of weighing using the calibration test

function or by using a certified test weight.

Control of this Calibration

Calibration using the internal weight is controlled by a

parameter in "Permission or prohibition". Refer to section

"8, Function Switch and Initialization".

n n M ( i

II U M M

----------- A--

-Calibration Using the internal Mass

0 ; Not used

1 :

Used

7, Calibration