Adapting to environment page 25 7, calibration – A&D GR Series GR-300 User Manual
Page 26
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".
step 6 Place the displayed target weight on the pan and
press the
| key. The balance displays the
measured target weight and displays it. Prevent
vibration and drafts from affecting the balance.
Step 7 Remove the weight from the pan after the balance
displays
L 11 U
innrtri
1 1
LUU.UU } i
9
r
1
cno
Step 8 If the "GLP output ( , n / - o ) " of the "Function Table
" is set to " /" or "P", "Calibration Test Report" is
output by the RS-232C interface.
Step 9 The balance will automatically return to the weigh
ing mode after calibration.
GLP output
Calibration test report
"end
nnnnn
UJJLiUU
9
Adapting to environment
page 25
7, Calibration