Measuring procedure

Step 1 Confirm thatthe balance is in the mode to measure

the weight of the sample in air. (" g" displayed and

processing indicator flashing)

Step 2 Confirm that the balance indicates zero. If it does

Step 3

not indicate zero, press the RE-ZERO key to reset

the displayed value to zero.

Place the sample on the upper pan (in air). When

the value displayed on the balance becomes stable,

Note

Step 4

press the [RANGE key to confirm the value (the

weight of sample in air). The balance enters the

mode to measure the weight of sample in liquid (

" g " displayed and processing indicator illuminat­
ing).

If a negative value or E (error) is displayed, the

RANGE key is disabled.

Place the sample on the lower pan (in liquid). When

the value displayed on the balance becomes stable.

press the [RANGE key to confirm the value (the

weight of sample in liquid). The balance enters the

mode to display the density (" g " not illuminated).

Note

Step 5

If E (error) is displayed, the [

range

! key is

disabled.

To output or save the density, press the [PRINT

key. The unit for outputting the density is " DS ".

To measure the density of another sample, press

the [RANGE key to return to the mode to measure
the weight in air and repeat the procedure de­

scribed above.

Note : If the liquid temperature or the type of liquid is

changed during measurement, reset the value
of the liquid density as necessary. For details,

see " Setting the density of a liquid ".