Calibration report using an external weight – A&D GR Series GR-300 User Manual

Page 44

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

Calibration Report using an external weight

Key operation

Step 1 Press and hold the

i

CAL| key until displaying

r

n I

L I I L

0

U c

Release the key.

n

u

Step 2 The balance displays

\ [ h ! i

□ If you want to change the calibration weight value, proceed to step 3.

□ If you use the stored calibration weight value in the balance, proceed to step 4.

Step 3 Press the [RANGE key and adjust calibration weight using the following keys.

RE-ZERO I key The key to set the value of the digit selected.

RANGE key The key to select the digit to change value.

№INT

I

key

The key to store a new weight value and return to step 2.

CAlj key

The key to cancel this change and return to step 2.

Step 4 Press the PRINT key. The zero point is measured and this value is

displayed.

Step 5 Place the calibration weight on the pan and press the PRINT key. The

weight is measured and this value is displayed.

Step 6 Remove the weight after [£noJ is displayed.

Step 7 If the calibration report is output

is output.

r /

n

LI L I

is displayed and the GLP data

Step 8 The balance returns to the weighing mode automatically.

AD-8121 format

_ r

I

I I I I LI I

Data format

i ii) i;i [..

(5 F?...... FF: i:i FFi

1-i

1 ;;F;;3 4F:F6 7TE;

^

iT3C:[:MEF-HiH

iFj !4 T FF:

F;

f

FF; .■■■■ FF: .:]■ ’ FFi FF^

1 -F FF: :;F: ;; 3- FF

f

F" i n

FF:FT,TFFF:FEiTT EF:F(F:::TT. F^

FFiT .. ¡JF’ T RFiT

— Factory —

Product —

Serial number

— ID number -
----- Date --------
----- Time---------

- Calibration type

Calibration weight

Signature

-- Space mark, ASCII 20H.

<TERM>

Terminator mark,

or S.

C

r

Carrige return mark, ASCII ODH

Lp Line feed mark, ASCII OAH

f

n I" O L

A & D<TERM>

MODEL

GR-200<T£flM>

S/N

12345678<TERM>

ID

ABCDEFGH<r£RM>

DATE<rtfiM>

<TERM>

TIME <TERM>

<TERM>

CALIBRATEDCEXT.)<T£RM>

CAL.WEIGHT<rERM>

-^200.0000 g<TERM>

SIGNATURE<rE,RM>
<TERM>
<TERM>

-----------------------------^TERM>

<TERM>

<TERM>

Example of GR-200

Functions

page 43

10, The ID Nunnber, GLP Report

Advertising
This manual is related to the following products: