Calibration Test

This function tests the balance for the accuracy using an external weight.
A weight which is used for the calibration test is called "the target weight".

The weight which you have is called "the external weight".

Operation

Step 1 Connect the AC adapter and warm up the balance

for at least one hour with nothing on the pan.

Step 2 Press and hold the CAL key until displaying

and then release the key.

LI U L

Press and hold the key

n

u

Step 3 The balance displays

If you want to change the target weight value,
proceed to step 4.

If you use the stored target weight value in the

balance, proceed to step 5.

Step 4 Press the j RANGE key and adjust the calibration

weight value using the following keys.

RE-ZERO [ key

RANGE key

PRINT key

CAL]key

The key to set the value of the

digit selected.
Thekeytoselectthedigittochange
value.
The key to store a new weight

value and return to step 3.
The key to cancel this change

and return to step 3.

Product

Usable weight

Adjustable range

GR-120

100g, 50g

GR-200

200g, 100g

-1-15.9 mg ~ -15.0 mg

GR-300

200g, 300g

GR-202

200g, 100g

Note

Digits cyclically change using the , RE-ZERO | key.
ex. 0mg^-h15mg-^-15mg-^0mg

Step 5 Confirm that there is nothing on the pan and press

the PRINT key. The balance measures the zero-

point and displays it. Prevent vibration and drafts

from affecting the balance.

7. Calibration