Advice, Control of automatic self calibration, Automatic self calibration – A&D GR Series GR-300 User Manual

Page 20: Caution

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

Automatic Self Calibration

This function automatically calibrates the balance, when the balance

detects an ambient temperature change.

When the balance detects a change in ambient tem­

perature, this indicator blinks and automatic self calibra­

tion is required. If the balance is not used for several

minutes with this indicator blinking, the balance pre­

forms automatic self calibration. The environment will

affect the time that the indicator blinks.

r i;ii

I. III.

This display means "the balance is measuring calibra­

tion data". Do not allow vibration or drafts to affect the

balance while this indicator is displayed.

Advice

You can use the balance while the indicator blinks. But, it is recommended

that to maintain the best accuracy, stop using the balance and confirm that

there is nothing on the pan when the indicator starts blinking.

Caution

□ Do not place anything on the weighing pan during automatic calibration.

Control of Automatic Self Calibration

Automatic self calibration is controlled by a parameter in "Permission or

prohibition". Refer to section "8. Function Switch and Initialization"

n n

II IJ

i 1 i

i 1 ?

-Automatic Self Calibration

0 : Not used

1 : Used

Adapting to environment

page 19