cknowledge Co

This is an explanation

o f t > - i d

of the function list

<AK> (06h) -—Acknowledge in ASCII code.

L I

n

u

In the Case of

h

□ The balance does not output <AK> code or the error code.

In the Case

o f F r i d

/

When the balance received a connmand requesting data and can not

process it, the balance transmits an error code ( EC, Exx).
When the balance is able to process a command requesting data, the

balance outputs the data.

When the balance receives a command to control the balance and can not

process it, the balance transmits an error code ( EC, Exx).

When the balance receives a command to control the balance and can

process it, the balance transmits <AK> (06h) code.

There are some commands that transmit plural <AK> (06h) code from the

balance. See "Command Examples"

CAL

command (Calibration command)

ON

command (ON command)

P

command (ON:OFF command)

R

command (RE-ZERO command)

TST

command (Calibration test)

When a communication error has occurred due to external noise, or a parity

error has occurred due to transmission error, the balance transmits an error
code. In this case, send the command once more.

Control using CTS and

This is an explanation

o f [ i - S

of the function list.

r;

U

In the Case of /_ t S

□ Regardless of whether the balance can receive a command or not, the

balance keeps the CTS line to Hi. The balance outputs data regardless of
condition of the RTS line.

In the Case

o f F h S I

□ The CTS line is kept HI normally. When the balance can not receive the next

command (ex. processing last command), the balance sets CTS line to LO.
The balance confirms the level of the RTS line when data can be output.

If the RTS level is HI, the balance outputs data. If the RTS level is LO, data
is not output (It cancels data output).