Step 6 Place the displayed calibration weight on the pan
and press the
key. The balance displays
the measured calibration weight. Prevent vibra
tion and drafts from affecting the balance.
Step 7 Remove the weight from the pan after the balance
displays |£nd !■
Step 8 If the "GLP output ( , n / - o ) " of the "Function Table
" is set to " /" or "P", "Calibration Test Report" is
output by the RS-232C interface.
Step 9 The balance will automatically return to the weigh
ing mode after calibration.
'rnn
1
u u
À....................
f i
:
[~ncl
GLP output
Calibration report
nnnnn
U.ULIUU
StepIO Test the accuracy of weighing using the calibra
tion test function with a certified test weight.
Control of this Calibration
Calibration using an external weight is controlled by
a parameter in "Permission or prohibition". Refer to
section "8. Function Switch and Initialization"