Step 6 Place the displayed calibration weight on the pan

and press the

PRINT

key. The balance displays

the measured calibration weight. Prevent vibra­
tion and drafts from affecting the balance.

Step 7 Remove the weight from the pan after the balance

displays |£nd !■

Step 8 If the "GLP output ( , n / - o ) " of the "Function Table

" is set to " /" or "P", "Calibration Test Report" is
output by the RS-232C interface.

Step 9 The balance will automatically return to the weigh­

ing mode after calibration.

'rnn

1

u u

À....................

f i

:

[~ncl

GLP output
Calibration report

nnnnn

U.ULIUU

StepIO Test the accuracy of weighing using the calibra­

tion test function with a certified test weight.

Control of this Calibration

Calibration using an external weight is controlled by

a parameter in "Permission or prohibition". Refer to
section "8. Function Switch and Initialization"

n n Î I i !

n ' U I I ! !

-Calibration Using the External Mass

u : Not used

I :

Used

Adapting to environnnent

page 23

7, Calibration