This function tests the balance accuracy using the internal weight.

Operation

Step 1 Connect the AC adapter and warm up the balance

for at least one hour with nothing on the pan.

o nnnnri

ULIUUU 9

Step 2 Press and hold the CAL key until display-

and then release the key.

Press and hold the key/C

ing

I I

L L in

f r a

r r

t L

i n

g

-Release the key

Step 3 The balance measures the zero point.

Prevent vibration and drafts to affect the balance.

Step 4 The measured zero point data is displayed.

Step 5 Ready for the internal weight measurement.

Step 6 The balance measures the internal weight.

Prevent vibration and drafts to affect the balance.

Step 7 The internal weight data is displayed.

Step 8 The balance informs you when the calibration test

is finished. If the "CLP output ( ,n/-o)" of fh®
"Function Table" is set to " /" or

" P " ,

"Calibration

Test Report" is output by the RS-232C interface.

Step 9 Thebalancewill automatically return to the weigh­

ing mode after the calibration test is finished.