Installation

13

Rinsing the Filter Pack cartridge

J

Let FILLING TANK mode run for a minimum of 2 hours with the Tank

Outlet Valve open and placed to the drain. This will completely rinse
the purification media inside the Filter Pack cartridge.

NOTE: for sensitive applications, it is recommended to leave the system

in FILLING TANK overnight to ensure complete rinsing of the RO

membrane.

Close the Tank Outlet Valve and connect it back to the system if

needed. The tank will start to fill up with water. It could take
approximately 1.5 to 2 hours to fill the tank. When it is full, the
Display will indicate a full tank by illuminating the symbols
representing the Tank Level

(see Section Understanding the Display)

.

The Vent Filter needs to be installed. See Section Installation of the

Vent Filter.