How to calibrate the tank level (c04), Ion how to calibrate the tank level (c04) – Labconco WaterPr BT Systems 90151 Series User Manual

Page 52

Advertising
background image

Troubleshooting

44

How to Calibrate the Tank Level (C04)

Before calibrating the tank level, the tank needs to be filled to the 100% level or TANK FULL.

1

Press the Main and “+“ Buttons together to
enter the menu. The Display will show C01.

2

Press the Main Button 3 times. The Display will
show C04. The highest Tank Level display will
be blinking. This means that the highest water
level is ready to be calibrated.

3

Press the “+” and “-“ Buttons together to enter
the highest water level. The Display will show
the lowest Tank Level (red level) display
blinking.

NOTE: The highest and lowest tank level

must be calibrated together or else the tank

level is not calibrated.

Continued on next page