Labconco WaterPr BT Systems 90151 Series User Manual

Page 60

Advertising
background image

Troubleshooting

52

Sanitizing the Tank

C

Open the front cover (this may require to remove the final filter) to go
into STANDBY mode (C).

D

Remove the Vent Filter and install the clear elbow fitting from the

Sanitization Kit (D).

E

Introduce the male connector of the clear tubing firmly into the

elbow fitting (E).

F

Inject 200 ml (1 ml = 1 cc) of Hydrogen Peroxide solution (30%) into

the 6 Liter tank via the clear tubing (F).

Rinse the air vent port with 200 ml of purified water.

Continued on next page

Advertising
This manual is related to the following products: