Labconco WaterPr BT Systems 90151 Series User Manual

Page 56

Advertising
background image

Troubleshooting

48

Sanitizing the System and the Tank

C

Open the front cover (this may require removal of the final filter) to let
the system go into STANDBY mode (C).

D

Remove the Vent Filter and install the clear elbow fitting from the

Sanitization Kit (D).

E

Introduce the male connector of the clear tubing firmly into the

elbow fitting (E).

F

Inject 200 ml (1 ml = 1 cc) of Hydrogen Peroxide solution (30%) into

the 6 Liter tank via the clear tubing (F).

Rinse the air vent port with 200 ml of purified water.

Continued on next page