Using the WaterPro® BT System

28

Operating Modes

Standby

STANDBY mode is displayed when the front
cover is removed. The system will
depressurize during which STANDBY will be
blinking on the Display for 10 seconds. All
system operations are disabled. STANDBY
mode is selected before attempting
maintenance on the system.

Flush

FLU (FLUSH) mode is displayed for
15 minutes after a new Filter Pack
cartridge has been installed and the front
cover has been closed. FLUSH mode allows
tap water to enter and rinse the new Filter
Pack cartridge. The Tank Outlet Valve must
be opened during the FLUSH cycle to keep
the tank from filling until the Filter Pack
cartridge is rinsed.

FLUSH mode is stopped if the front cover is
removed to go into STANDBY mode. When
the front cover is closed, FLUSH mode
resumes from the last remaining time on
the Display.

If the system is powered OFF during FLUSH
mode and powered back ON, then a new
FLUSH cycle will start.

The system will have a 2 minute FAST FLUSH
cycle every 24 hours of PRE OPERATE mode.