Troubleshooting – Labconco WaterPr BT Systems 90151 Series User Manual

Page 69

Advertising
background image

Troubleshooting

61

Problem

Possible causes

What to do

The Display screen is blank.

The power cord is not plugged in.

No source of electrical power.

Main Power Fuse is blown.

Check that the power cord is

plugged in.

Check the source of electrical

power.

Contact your technical support.

In FILLING TANK mode, the
Tank Level display is not

rising.

(The Tank Level display

should be rising steadily.)

The Tank Outlet Valve is open. The

water in the tank is diverted into the

drain. No water is kept in the tank.

The tank level is not calibrated

properly.

The RO membrane is dirty or

clogged.

Close the Tank Outlet Valve.

Check the Feedwater source.

See

Section How to Calibrate the

Tank Level (C04).

Replace the Filter Pack cartridge.

Low water flow or no water
flow when the Dispensing
Button is pressed.

The water level is less than 10%.

No dispensing is allowed.

The Tank Outlet Valve is open. The

tank is not filling up with water.

.22 um Filter is air locked.

Final Filter is clogged.

Allow the system to go into

FILLING TANK mode until there is

enough water available in the tank,
viewed by the Tank Level display.

Close the Tank Outlet Valve.

Vent all air from the .22um Filter.

Replace the Final Filter, see

Section

How to Replace the Final Filter.

DISPENSING mode
continuously runs even

when the Tank Level display
is below 10%.

The water level is not calibrated.

The system does not detect the
minimum level to disable
DISPENSING mode.

Calibrate the tank level. See

Section How to Calibrate the Tank

Level (C04).

AUTO-DISPENSING mode is
not accurate.

The Product Water flowrate is not

calibrated.

The Final Filter is clogged or

airlocked.

Calibrate the Product Water

flowrate. See

Section How to

Calibrate the Flowrate from the

WaterPro® BT System

(F02)

.

Replace the Final Filter.

The last RO Permeate
conductivity value is not
displayed when you press
the “+” button.

The measurement was not made

while dispensing water during
FILLING TANK mode.

The value is out of measurement

range.

Dispense water during FILLING

TANK mode to start a RO Permeate

conductivity reading again.