Using the WaterPro® BT System

31

How to view the Product Resistivity and Temperature in Filling Tank mode or in Pre Operate

mode

Press:

The Display will show the last Product
resistivity and temperature values
measured during DISPENSING mode or
during RECIRCULATION. The values are
displayed for 5 seconds.

NOTE: The Product resistivity and

temperature are displayed automatically

during DISPENSING mode or during

RECIRCULATION.

How to view the RO Permeate Conductivity in Filling Tank mode or in Pre Operate mode

Press:

The Display will show the RO Permeate
Water conductivity value that was
measured during Filling Tank mode. The
displayed value is not a real-time value.
It is the last permeate conductivity value
memorized after the system was in
Dispensing Mode. The value is displayed
for 5 seconds.

NOTE: The units are displayed in µS/cm

only.

How to recirculate water manually before dispensing

This option is used to enhance the quality of the Product Water before dispensing water. The Distribution Pump
will turn on and water will recirculate for up to 3 minutes. An auto-recirculation occurs for 3 minutes every
2 hours.

Press:

The Display will show the Product
resistivity and temperature.

NOTE: To dispense water, press the

Dispensing Button during

recirculation. Recirculation will stop

and water will be dispensed.