Installation

19

4

Measure the total volume of water (in
Liters) dispensed from the system using a
1 L graduated cylinder.

5

Press the “+” or “-“ Button to match the
volume Display to the volume measured.

6

To exit the menu, press and hold the Main
Button for 2 seconds.

To display the next menu option, press the
Main Button once.

The Product Water flowrate is now
calibrated.