Troubleshooting

55

How to Replace the UV Lamp (UV System only)

The red UV Lamp Alarm will be blinking on the Display when it is time to exchange the UV Lamp. The message is
shown when the UV Timer has reached 0 days

(

see

Section How to View or Reset the UV Lamp Timer (C05)).

HAZARD

No electrical power should be going to the system at this time. Accidental
exposure to ultraviolet light can cause damage to the eyes and skin.

A

Open the front cover (this may require removal of the final filter) to go
into STANDBY mode (A).

B

Unplug the electrical cord to power OFF the system.

C

Remove the Final Filter.

Remove the Filter Pack cartridge.

Advertising