Installation

20

How to Show Resistivity or Conductivity Units (C01)

With Temperature Compensated or non Temperature Compensated values

Temperature compensation is a way of ‘standardizing’ Resistivity or Conductivity to measurements that would be
seen if the water temperature was 25 °C.

1

Press the Main and “+“ Buttons together
to enter the menu. The Display will show
C01 and the units chosen. The following
Display shows Temperature Compensated
Resistivity Units: MΩ.cm @25 °C.

2

Press the “+” or “-“ Button to select
Non Temperature Compensated
Resistivity Units: MΩ.cm.

3

Press the “+” or “-“ Button to select
Temperature Compensated Conductivity
Units: µS/ cm @25 °C.