Using the WaterPro® BT System

29

Filling Tank

FILLING TANK mode is displayed when the
tank is being filled with RO water until the
100% Tank Level display. FILLING TANK
mode is launched automatically when the
Tank Level display is below the 60% level
or after a FLUSH cycle has been completed.

FILLING TANK mode

Water can be dispensed or can be
periodically recirculated during FILLING
TANK mode if the Tank Level display is
above the 10% level.

Dispensing or Recirculation during

FILLING TANK mode

Pre Operate

PRE OPERATE mode is displayed when the
system is not dispensing water and not in
FILLING TANK mode. The Tank Level display
is between the 60% level and TANK FULL
level.

The system will have a 3-minute auto-
recirculation every 2 hours in PRE OPERATE
mode. The Distribution Pump turns on. This
will enhance the quality of the Product
Water delivered from the system.
Recirculation can also be activated
manually for up to 3 minutes.

The resistivity and temperature of the
Product Water is displayed during
recirculation. The resistivity and
temperature remain displayed for up to
10 seconds after recirculation is finished.