Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Exhaust and Blower Requirements

The XPert Nano Enclosure uses an integral motorized impeller to draw room air
past the operator and through the enclosure. This contaminated air is then pushed
through the ULPA filter. The ULPA-filtered exhaust air is then forced out the top
of the enclosure.

The ULPA-filtered exhaust air can be recirculated into the laboratory or exhausted
outside with the addition of the Canopy Exhaust Connection and remote blower
listed in Chapter 7. Electrical connections are covered in Chapter 2 for this
configuration. Consult your local regulations or guidelines for ventilation of the
filtered exhaust in nanoparticle applications.

If external exhaust is required, data for the exhaust volume, noise pressure and
enclosure static pressure loss is listed for each filtered enclosure model at face
velocities of 60, 80, and 100 fpm. Weighing and handling nano materials is best done
with lower flow rates (60-70 fpm or 0.31-0.36 m/s).

Enclosure

Width and

Sash

Opening

Face Velocity

(fpm)

Exhaust

Volume

(CFM)

XPert Station

Initial Static

Pressure Loss

with HEPA

filter (in w.g.)

XPert

System

Noise

Pressure db

(A)

Xpert System

Max.

Equivalent

Resistance (Ft)

Xpert System

Max. External

Static Pressure

2'

9.4" Sash

60
80

100

85

115
145

.35"- .41"

.48" - .57"
.62" - .72"

48-53
50-56
53-58

400
200

80

.20"
.18"
.13"

3'

9.4" Sash

60
80

100

130
175
220

.37" - .43"
.53" - .62"
.68" - .79"

48-53
53-57
58-61

130

50
25

.15"
.11"
.08"

4'

9.4" Sash

60
80

100

175
235
295

.40" - .46"
.57" - .66"

.76" - .87".

49-54
58-61
62-66

100

50
25

.20"
.18"
.13"

5'

12" Sash

60
80

100

290
385
480

.38" - .46"
.55" - .67"
.72" - .85"

52-56
58-63
63-67

450
230
120

.22"
.20"
.15"

6'

12" Sash

60
80

100

350
465
580

.46" - .56"
.67" - .81"

.87" – 1.03"

55-57
61-63
67-69

300
150

70

.20"
.18"
.13"

Proper blower selection can be determined from these exhaust requirements and the
total system static pressure loss. For outside exhaust, the enclosure must be
connected to either a dedicated blower or a house exhaust system. Labconco offers
accessory remote blowers listed in Chapter 7. Contact Labconco for blower sizing
assistance.

If the enclosure is connected directly to a house exhaust system, an
adjustable damper (or valve) must be installed to control the airflow
properly. This is equally important when a house exhaust system is
controlling multiple filtered enclosures. See Chapter 7 for accessory
adjustable damper ordering information.

!